Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 107
Preparation for Accurate Measurement (Executing Correction)
Obtaining Correction Data
Description of softkeys
To enable/disable or otherwise control the behavior of load correction, use the
following softkeys:
Selecting single/multiple correction mode
Functional description
E4981A can store up to 256 sets of OPEN/SHORT/LOAD correction data. In
addition, it can store one set of the standard’s reference value data at a
specified frequency point. In multiple correction mode, you can switch among
up to 256 data sets to carry out correction.
In multiple correction mode, the channel number for which correction data is
selected.
To set up single/multiple correction mode:
Step 1. Press [Meas Setup].
Step 2. Press the CORRECTION softkey.
Step 3. Use the cursor keys to select the MULTI field.
Step 4. Use the following softkeys:
Step 5. Use the cursor keys to select the CH field.
Softkey Description
ON Enables load correction.
OFF Disables load correction.
MEAS LOAD Starts load correction.
The correction features default to single correction mode.
Softkey Description
ON
Turns ON the multiple correction mode.
OFF
Turns OFF the multiple correction mode.
For more information on measurement parameters, see “Setting up
measurement parameters” on page 40.