172 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Specifications and Supplemental Information
Basic Specifications
10-
100 µF 68 µF - 150 µF 0 F - 150 µF
220 µF 150 µF - 330 µF 0 F - 330 µF
470 µF 330 µF - 680 µF 0 F - 680 µF
1 mF 680 µF - 2 mF 0 F - 2 mF
Table 10-3 Measurable capacitance ranges when measurement frequency is 1 kHz
Measurement value
Recommended
measurement range
Significant
measurement range
100 pF 0 F - 150 pF 0 F - 150 pF
220 pF 150 pF - 330 pF 0 F - 330 pF
470 pF 330 pF - 680 pF 0 F - 680 pF
1 nF 680 pF - 1.5 nF 0 F - 1.5 nF
2.2 nF 1.5 nF - 3.3 nF 0 F - 3.3 nF
4.7 nF 3.3 nF - 6.8 nF 0 F - 6.8 nF
10 nF 6.8 nF - 15 nF 0 F - 15 nF
22 nF 15 nF - 33 nF 0 F - 33 nF
47 nF 33 nF - 68 nF 0 F - 68 nF
100 nF 68 nF - 150 nF 0 F - 150 nF
220 nF 150 nF - 330 nF 0 F - 330 nF
470 nF 330 nF - 680 nF 0 F - 680 nF
1 µF 680 nF - 1.5 µF 0 F - 1.5 µF
2.2 µF 1.5 µF - 3.3 µF 0 F - 3.3 µF
4.7 µF 3.3 µF - 6.8 µF 0 F - 6.8 µF
10 µF 6.8 µF - 15 µF 0 F - 15 µF
22 µF 15 µF - 33 µF 0 F - 33 µF
47 µF 33 µF - 68 µF 0 F - 68 µF
100 µF 68 µF - 200 µF 0 F - 200 µF
Table 10-2 Measurable capacitance ranges when measurement frequency is 120 Hz
Measurement value
Recommended
measurement range
Significant
measurement range