Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 125
Sorting Based on Measured Results (Comparator Function)
Setting Up Sorting Judgment Conditions
Relationship between the limit range designation method and the
setup value
Table 7-1 and Table 7-2 compare the setup values between the limit
designation methods when setting up the limit ranges shown in Figure 7-1 and
Figure 7-2, respectively.
Figure 7-1 Example of limit range settings (case 1)
Table 7-1 Lower and upper limit values of limit ranges for Figure 7-1 (comparison of
modes)
Absolute mode
Tolerance mode (reference value: 1 nF)
Absolute Percent
Lower
limit
value
Upper
limit
value
Lower
limit
value
Upper
limit
value
Lower
limit
value
Upper
limit
value
BIN1 0.98 nF 0.99 nF –0.02 nF –0.01 nF –2% –1%
BIN2 0.99 nF 1.01 nF –0.01 nF 0.01 nF –1% 1%
BIN3 1.01 nF 1.02 nF 0.01 nF 0.02 nF 1% 2%