124 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Sorting Based on Measured Results (Comparator Function)
Setting Up Sorting Judgment Conditions
7-
Setup procedure for the limit range designation method and reference
value
Step 1. Press [Meas Setup].
Step 2. Press the LIMIT TABLE softkey.
Step 3. Use the cursor keys to select the MODE field.
Step 4. Use the following softkeys:
Step 5. Use the cursor keys to select the NOM field, to set the reference value for
absolute tolerance mode or percent tolerance mode.
Step 6. Enter the nominal value using the softkeys or entry keys. If you use the entry
keys to enter the value, the softkey labels change to unit labels (p, n, u, m, x1).
Mode
Off Absolute mode
ΔABS
Tolerance mode
Absolute tolerance mode
Δ% Percent tolerance mode
Softkey Function
OFF Switches the comparator into absolute mode.
ΔABS Switches the comparator into tolerance mode based on absolute value.
Δ% Switches the comparator into tolerance mode based on deviation
percentages.
Softkey Function
INCR++ Increments the nominal value in steps of your selected number (1, 2, 5,
10, 20, 50, 100, 200, 500)
INCR+ Increments the nominal value in steps of 1.
DECR- Decrements the nominal value in steps of 1.
DECR-- Decrements the nominal value in steps of your selected number (1, 2, 5,
10, 20, 50, 100, 200, 500)