Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 109
Preparation for Accurate Measurement (Executing Correction)
Obtaining Correction Data
Setup procedure
Step 1. Press [Meas Setup].
Step 2. Press the CORRECTION softkey.
Step 3. Use the cursor keys to select the OFFSET field.
Step 4. Use the following softkeys:
Step 5. Use the cursor keys to select the OFFSET A/B field
Step 6. Use the numeric entry keys to enter the correction value for primary and
secondary parameter.
The entered values are set up as the data for the OFFSET correction for the
measurement frequency at the time of data entry.
Softkey Description
ON Turns ON the OFFSET correction function
OFF Turns OFF the OFFSET correction function