150 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Using Handler Interface
Electrical Characteristics
8-
Figure 8-6 Circuit diagram of the handler interface input signals
Table 8-7 Selection of handler input trigger voltage
:SYSTem:HANDler:TRIGger:VOLTage <para>
5 V ~ para | 9 V 9 V ~ para | 15 V 15 V ~ para ≤ 24 V
S1 ON OFF OFF
S2 OFF ON OFF
S3 OFF OFF ON
The S1, S2, S3 switches belongs to electrical switches that can be control
by SCPi command :SYSTem:HANDler:TRIGger:VOLTage <para>
e4981aoe0001
2.27k㱅 1k㱅
S2 S1
1.6K㱅
S3
3.68k㱅
EXT_ DCV2
EXT_TRIG
/KEY_LOCK
Handler
interface
connector