184 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Specifications and Supplemental Information
Basic Specifications
10-
Figure 10-6 Accuracy of Cp and Cs when measurement frequency is 1 kHz
(measurement range: 100 pF to 10 µF / measurement signal level: 1 V)
Figure 10-7 Accuracy of D when measurement frequency is 1 kHz
(measurement range: 22 µF to 100 µF / measurement signal level: 1 V)
e4981cue0030
Accuracy (Cp, Cs) [%]
0
0.05
0.1
0.15
0.2
0.25
Measurement Value (Cp, Cs) [F]
1.0E-11 1.0E-10 1.0E-9 1.0E-8 1.0E-7 1.0E-6 1.0E-5 1.0E-4
N=8
N=4
N=1
e4981cue0033
Accuracy (D)
0.0042
0.0043
0.0044
0.0045
0.0046
0.0047
0.0048
0.0049
0.005
Measurement Value (Cp, Cs) [F]
0.00001 0.0001 0.001
N=8
N=4
N=1