Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 193
Specifications and Supplemental Information
Supplemental Information
Signal Level Compensation (SLC) function
SLC function compensates the voltage drop by resistance inside the E4981A
and the extension cable under the following frequencies and ranges:
Measurement Cable: 16048A or 16048D
Operating measurement range setting for 1 m or 2 m cable length is:
When the measurement frequency is 120 Hz: 220 µF range, 470 µF range,
1 mF range
When the measurement frequency is 1 kHz: 22 µF range, 47 µF range,
100 µF range