196 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Specifications and Supplemental Information
Supplemental Information
10-
Measurement circuit protection
The maximum discharge withstand voltage, where the internal circuit remains
protected if a charged capacitor is connected to the UNKNOWN terminal, is
illustrated below.
Figure 10-12 Maximum discharge withstand voltage (typical)
Discharge capacitors before connecting them to the UNKNOWN terminal
or a test fixture.
Table 10-19 Maximum discharge withstand voltage (typical)
Maximum discharge withstand
voltage
Range of capacitance value C of DUT
1000 V C | 2 µF
V
C ~ 2 µF