240 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Error Messages
Error Messages (Alphabetical Order)
-
T
-124 Too many digits
The number of digits of the mantissa of the decimal value data element
exceeds 255 except for preceding 0s. (Refer to IEEE488.27.7.2.4.1.)
-211 Trigger ignored
A trigger command or trigger signal has been received and recognized by the
E4981A, but it is ignored due to the timing relationship with the E4981A. For
example, this happens when the E4981A’s trigger system is not in the Waiting
for Trigger state.
-214 Trigger deadlock
Indicates that the :READ? command was ignored because the trigger source
setting was MAN or BUS.
-223 Too much data
The received block, equation, or string type program data complies with the
standard, but the amount of data exceeds the limit that the E4981A can handle
due to memory or device-specific conditions related to memory.
U
-113 Undefined header
A header not defined for the E4981A has been received. For example, “*XYZ”,
which is not defined for the E4981A, is received.