46 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Getting Started
Learning Basic Measurement Procedure
3-
Checking data for OPEN correction
You can check the measured open admittance value (data for the OPEN
correction). The procedure is given below.
Step 1. Press the [Meas Setup] key.
Step 2. Press the CORRECTION softkey.
Step 3. Use the cursor keys to select the OPEN G-B field.
Figure 3-10 OPEN Correction Data
Step 4. Use the following softkeys:
Softkey Description
G-B Displays the measured G-B value on the screen.
Cp-G Displays the measured Cp-G value on the screen.