284 Keysight E5071C ENA Network Analyzers
Post-Repair Procedures
Post-Repair Procedures
7-
Table 7-1 Post-Repair Procedures for Analog Assemblies for Up to 8.5 GHz Options
Replaced
Assembly
or Part
Required Adjustments
Correction Constants
(CC)
Verification
PCI-DSP
Card (A51)
Perform the following
required adjustment.
- Write ID
- Save to Flash memory
Perform “To Execute the diagnostics
Test” on page 80
Perform the following performance test.
- AUX Input Test
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness
Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic Accuracy Test
- Uncorrected System Performance Test
Perform “To Check the External
Trigger Input” on page 99
Reference
Oven Board
(A5)
Perform the following
required adjustment.
- OCXO
Perform “To Execute the diagnostics
Test” on page 80
Perform the following performance test.
- Frequency Accuracy Test
Analog
Base
Module
(A6)
Perform the following
required adjustment.
- AUX Input
- Frequency Reference
Perform “To Execute the diagnostics
Test” on page 80
Perform the following performance test.
- AUX Input Test
- Frequency Accuracy Test