Keysight E5071C ENA Network Analyzers 285
Post-Repair Procedures
Post-Repair Procedures
Synthesizer
Module
(A1)
Perform the following
required adjustment.
- Synthesizer Gain
- Source Output Power
- Receiver Port
Characteristics
- Receiver Absolute Gain
Perform “To Execute the diagnostics
Test” on page 80
Perform the following performance test.
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness
Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic Accuracy test
- Uncorrected System Performance Test
Receiver
Module
(A2)
Perform the following
required adjustments.
- Source Output Power
- Receiver IF Range
- Receiver Port
Characteristics
- Receiver Absolute Gain
Perform “To Execute the diagnostics
Test” on page 80
Perform the following performance test.
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic accuracy test
- Uncorrected system performance
Table 7-1 Post-Repair Procedures for Analog Assemblies for Up to 8.5 GHz Options
Replaced
Assembly
or Part
Required Adjustments
Correction Constants
(CC)
Verification