EasyManua.ls Logo

Keysight E5071C ENA - Page 75

Keysight E5071C ENA
361 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight E5071C ENA Network Analyzers 75
Troubleshooting
Boot Process Troubleshooting
Step 4. Measurement view
The measurement view as shown in Figure 4-8 is displayed after the system
initialization is completed without problem.
Figure 4-8 Measurement view
Step 5. Power-on self test
The power-on self test is executed once automatically before the
measurement starts. While the power-on self test is in progress, "Power on
test" is displayed at the left in the instrument status bar. If the power-on test
fails, an error message is displayed there. For more details, refer to
“Troubleshooting Using Diagnostics Test” on page 76.