309 Cannot initiate, voltage and current in fixed mode
Cannot initiate transient generator. Either the voltage or current function is set to Fixed mode.
310 The command is not supported by this model
This instrument does not have either the hardware or options required to support this command.
315 Settings conflict error
A data element could not be programmed because of the present instrument state.
316 Mass storage error
The mass storage memory has been exceeded.
317 Invalid format
An invalid data format was found in the command string.
320 Firmware update error
This may be due to the instrument hardware not being able to support the firmware version.
324 Inconsistent arb settings
The arb settings are inconsistent; most likely a mismatch in the arb lengths.
327 Initiated with no sense function enabled
A measurement has been initiated without specifying the measurement (sense) function.
328 Too many measurement points
Too many measurement points have been specified.
331 Illegal parameter value
The parameter value is out of range or does not exist.
334 Table number is invalid
The table number must be either 1 or 2.
335 VMP must be less than VOC; VMP setting is <setting>, VOC setting is <setting>
The VMP curve parameter is equal to or greater than the VOC parameter
337 IMP must be less than or equal to ISC
The IMP curve parameter is greater than the ISC parameter
339 VMP and/or IMP too small
The VMP and/or IMP are less than the allowable range of values
Keysight MP4300 Series Operating and Service Guide 159
5 SCPI Programming Reference