Command Errors (these errors set Standard Event Status register bit #5)
-100 Command error
Generic syntax error.
-101 Invalid character
An invalid character was found in the command string.
-102 Syntax error
Invalid syntax was found in the command string. Check for blank spaces.
-103 Invalid separator
An invalid separator was found in the command string. Check for proper use of , ; :
-104 Data type error
A different data type than the one allowed was found in the command string.
-105 GET not allowed
A group execute trigger is not allowed in a command string.
-108 Parameter not allowed
More parameters were received than were expected.
-109 Missing parameter
Fewer parameters were received than were expected.
-110 Command header error
An error was detected in the header.
-111 Header separator error
A character that was not a valid header separator was found in the command string.
-112 Program mnemonic too long
The header contains more than 12 characters.
-113 Undefined header
A command was received that was not valid for this instrument.
-114 Header suffix out of range
The value of the numeric suffix is not valid.
-120 Numeric data error
Generic numeric data error.
5 SCPI Programming Reference
160 Keysight MP4300 Series Operating and Service Guide