SENT Triggering and Serial Decode 32
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 509
 
• Slow Channel Message ID — finds Slow Channel Message IDs that match the 
value entered using additional softkeys.
• Slow Channel Message ID & Data — finds Slow Channel Message IDs and Data 
that match the values entered using additional softkeys.
• All CRC Errors — finds any CRC error, Fast or Slow.
• Pulse Period Error — finds where a nibble is either too wide or too narrow (for 
example, data nibble < 12 (11.5) or > 27 (27.5) ticks wide). Sync, S&C, data, 
or checksum pulse periods are checked.
For more information on searching data, see “Searching Lister Data" on page 152.
For more information on using the [Navigate] key and controls, see “Navigating the 
Time Base" on page 75.