Searching for MIL-STD-1553 Data in the Lister
The oscilloscope's search capability lets you search for (and mark) certain types of 
MIL-STD-1553 data in the Lister. You can use the [Navigate] key and controls to 
navigate through the marked rows.
1 With MIL-STD-1553 selected as the serial decode mode, press [Search].
2 In the Search Menu, press the Search softkey; then, turn the Entry knob to select 
the serial slot (Serial 1 or Serial 2) on which the MIL-STD-1553 signal is being 
decoded.
3 Press Search; then, select from these options:
• Data Word Start — finds the start of a Data word (at the end of a valid Data 
Sync pulse).
• Command/Status Word Start — finds the start of Command/Status word (at the 
end of a valid C/S Sync pulse).
• Remote Terminal Address — finds the Command/Status word whose RTA 
matches the specified value. The value is specified in hex.
When this option is selected, the RTA softkey becomes available and lets you 
select the hex Remote Terminal Address value to find.
• Remote Terminal Address + 11 Bits — finds the RTA and the remaining 11 bits 
that match the specified criteria.
When this option is selected, these softkeys become available:
• The RTA softkey lets you select the hex Remote Terminal Address value.
• The Bit Time softkey lets you select the bit time position.
• The 0 1 X softkey lets you set the bit time position value as a 1, 0, or X 
(don't care).
• Parity Error — finds (odd) parity bits that are incorrect for the data in the word.
• Sync Error — finds invalid Sync pulses.
• Manchester Error — finds Manchester encoding errors.
For more information on searching data, see “Searching Lister Data" on page 152.
For more information on using the [Navigate] key and controls, see “Navigating the 
Time Base" on page 75.