4 System
4.6 Alignments
Query Temperature of Last Align Now RF
Remote
Command
:CALibration:TEMPerature:LRF?
Example
:CAL:TEMP:LRF?
Notes Value is in degrees Centigrade at which the last successful Align Now RF was executed, either
individually or as a component of Align Now All
State Saved No
Query Time since Last Align IF
Remote
Command
:CALibration:TIME:LIF?
Example
:CAL:TIME:LIF?
Notes Value is the elapsed time, in seconds, since the last successful Align IF was executed
State Saved No
Query Temperature of Last Align IF
Remote
Command
:CALibration:TEMPerature:LIF?
Example
:CAL:TEMP:LIF?
Notes Value is in degrees Centigrade at which the last successful Align IF was executed
State Saved No
Query Time since Last Characterize Preselector
Remote Command
:CALibration:TIME:LPReselector?
Example
:CAL:TIME:LPR?
Notes Value is the date and time the last successful Characterize Preselector was executed. The date is
separated from the time by a space character
Returns “” if no Characterize Preselector has ever been performed on the instrument
Dependencies In models that do not include preselectors, this command is not enabled and any attempt to set or
query yields an error
State Saved No
Query Temperature of Last Characterize Preselector
Remote Command
:CALibration:TEMPerature:LPReselector?
Example
:CAL:TEMP:LPR?
Notes Value is in degrees Centigrade at which the last successful Characterize Preselector was executed
Short Range Comms & IoT Mode User's &Programmer's Reference 1836