4 System
4.6 Alignments
Dependencies In models that do not include preselectors, this command is not enabled and any attempt to set or
query yields an error
State Saved No
Query Time since Auto Align Off
Remote
Command
:CALibration:AUTO:TIME:OFF?
Example
:CAL:AUTO:TIME:OFF?
Notes Value is the elapsed time, in seconds, since Auto Align has been set to Off or Off with Alert. The value is
0 if Auto Align is ALL or NORF
State Saved No
Query Time since Last Align Now 20 Hz 30 MHz
Remote
Command
:CALibration:TIME:RFPSelector:LCONducted?
Example
:CAL:TIME:RFPS:LCON?
Notes Values are the date and time the last successful Align Now, 20Hz 30MHz was executed. The date is
separated from the time by a semi-colon character
State Saved No
Query Temperature of Last Align Now 20 Hz 30 MHz
Remote
Command
:CALibration:TEMPerature:RFPSelector:LCONducted?
Example
:CAL:TEMP:RFPS:LCON?
Notes Value is in degrees Centigrade at which the last successful Align Now, 20Hz 30MHz was executed
State Saved No
Query Time since Last Align Now 30 MHz 3.6 GHz
Remote
Command
:CALibration:TIME:RFPSelector:LRADiated?
Example
:CAL:TIME:RFPS:LRAD?
Notes Value is the date and time the last successful Align Now, 30MHz 3.6GHz was executed. The date is
separated from the time by a semi-colon character
State Saved No
1837 Short Range Comms & IoT Mode User's &Programmer's Reference