EasyManua.ls Logo

Kodak M35 - Recirculation Pump

Kodak M35
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Recirculation Pump
Replacement of the Recirculation Pump [10] Connect the water supply.
[11] Remove the PINCH CLAMPS.
[12] Fill the 3 TANKS with water.
Dangerous voltage.
[13] Connect the main power.
[1] Disconnect the main power.
[2] Disconnect the water supply.
[3] Remove the TOP COVER and the 2 SIDE
Never operate the RECIRCULATION
PANELS.
PUMPS if the TANKS are empty.
[4] Drain the 3 TANKS.
[14] Actuate the RECIRCULATION PUMPS.
[5] Use PINCH CLAMPS on the tubing to prevent
[15] Check for agitation on the surface of the
leakage.
solutions.
[6] Disconnect the tubing to the
[16] Check for leakage.
RECIRCULATION PUMP.
[17] Drain the water thoroughly from the developer
[7] Remove the MOUNTING SCREWS.
and fixer TANKS and fill the TANKS with
[8] Remove the RECIRCULATION PUMP.
developer and fixer.
[9] Reverse the above procedure to install a new
RECIRCULATION PUMP.
H112_0071BA
H112_0071BCE
MOUNTING
SCREW
RECIRCULATION PUMP
Replacement of the Recirculation Pump
2-34 981777