Recirculation Pump
Replacement of the Recirculation Pump [10] Connect the water supply.
[11] Remove the PINCH CLAMPS.
[12] Fill the 3 TANKS with water.
Dangerous voltage.
[13] Connect the main power.
[1] Disconnect the main power.
[2] Disconnect the water supply.
[3] Remove the TOP COVER and the 2 SIDE
Never operate the RECIRCULATION
PANELS.
PUMPS if the TANKS are empty.
[4] Drain the 3 TANKS.
[14] Actuate the RECIRCULATION PUMPS.
[5] Use PINCH CLAMPS on the tubing to prevent
[15] Check for agitation on the surface of the
leakage.
solutions.
[6] Disconnect the tubing to the
[16] Check for leakage.
RECIRCULATION PUMP.
[17] Drain the water thoroughly from the developer
[7] Remove the MOUNTING SCREWS.
and fixer TANKS and fill the TANKS with
[8] Remove the RECIRCULATION PUMP.
developer and fixer.
[9] Reverse the above procedure to install a new
RECIRCULATION PUMP.
H112_0071BA
H112_0071BCE
MOUNTING
SCREW
RECIRCULATION PUMP
Replacement of the Recirculation Pump