Adjustment of the Replenishment Flow Rates
in M35A Processors with Serial No. 10,000 and Above or
in M35 Processors with Serial No. 35,000 and Above
[1] Remove the TOP COVER.
[2] Lift the top DETECTOR ROLLER of the
DETECTOR CROSSOVER ASSEMBLY.
[3] Check that the replenisher solutions flow
freely through the TUBING along the drive
side of the processor.
[4] Press the METAL LATCH on the red QUICK
DISCONNECT to disconnect the developer
TUBING.
[5] Pull the TUBING slightly by rotating the
TUBING over the edge of the frame.
[6] Insert the TUBING into a GRADUATE.
[7] Lift the top DETECTOR ROLLER of the
H112_0114AC
H112_0114ACB
GRADUATE
DISCONNECT
QUICK
TUBING
LATCH
METAL
DETECTOR CROSSOVER ASSEMBLY for
the correct time, 28 or 34 seconds. See the
Adjusting the Replenishment Flow Rate
table on the next page.
in M35A Processors with Serial No. 10,000 and
Above or in M35 Processors with Serial No.
NOTE
35,000 and Above
The REPLENISHMENT PUMP will
operate 3 seconds after you release
the DETECTOR ROLLER.
[8] Check that the amount of developer in the GRADUATE is the same as in the table.
[9] Adjust the REPLENISHMENT PUMP if necessary. See page 2-39 for the adjustment procedure.
[10] Do steps 7 - 9 again if necessary.
[11] Connect the QUICK DISCONNECT.
[12] Do this procedure again to check the flow rate of the fixer.