Chemical Replenishment
[1] Check the replenishment flow rates by doing the procedure on page 2-40 or page 2-42.
[2] Check the STRAINERS for debris.
(a) Place clamps on the tubing to stop the replenishment flow.
(b) Disassemble the STRAINER ASSEMBLIES.
(c) Use a brush and warm water to clean the STRAINER ASSEMBLIES and to remove dirt and
chemical deposits from the SCREENS.
H112_0092BA
H112_0092BCA
SCREEN
STRAINER ASSEMBLY
Cleaning the Replenishment Strainer Assemblies
981777 3-15