1. Transport Failure
2. Surface Artifacts
3. Abnormal Film Densities
4. Wet Films
5. Low Solution Levels
6. Overlapping of Films
123456
• Check the time delay. For all transport speeds, the buzzer should sound
once the trailing edge of the film has advanced 75 mm (3 in.) into the
processor.
••• •Check that the tank solution levels are at the overflow weirs.
•• Adjust the guide shoes correctly.
NOTE: Most often in the developer turnaround. Adjust the tips of the
guide shoes in the exit turnaround as close as possible to the rack
roller above them.
• Check for solution in the replenisher tanks. Fill if necessary.
NOTE: Do not mix more than a 2-week supply of developer replenisher.
• Check that the correct bulb is in the safelight.
981777 4-3