KurzB‐SeriesOperationsGuide 1–13
B‐SeriesFlowMeterModes & Menus
16 REMOTECF RemoteCorrectionFactorData
RemoteCF
ON OFF
Enter#ofRemoteCFDataPointsn
EnterExtInput.RemCF‐D1
EnterExtInput.RemCF‐X1
:
EnterExtInput.RemCF‐Dn
EnterExtInput.RemCF‐Xn
Thisallowsyouto
correctthemeter
readinginreference
tootherdata.
17 CALCURVE GasCalibrationCurveorGasConcentration
Selection(optional)
CurveSel
Mode
MANUALSELECT
EXTINPU TLEV
IfMANUALSELECT
CalCurve#
Theselection
dependsonthe
meterconfiguration.
18 DATALOG DataLoggin gSetup
EnableDataLog
ON OFF
IfON
LogIntervalsec
Setsupperiodic
loggingofmeasured
flowandtotalflow
totheUSBport.
19 MODBUSCOM ModbusCommunicationSetup
DevModbusAddr
Modbus
Mode
MODBUSRTU MODBUSASCII
ModbusBaudRate
9600 38400
14400 57600
19200
RegisterOrder
BYTE#1234 BYTE#3412
Allowsyoutosetup
upModbus
communications
andspecifyeither
RTUorASCII
commands.The
baudratewill
typicallybe9600.
20 EXTAINPUT ExternalInputSetup
ExtInputUsage
CALDATASW PIDEXT.REF(optional)
VMREFERENCE REMOTECF
PURGECOMMAND
(optional)
VMREFERENCE
ScaleUnit
InpValAt4mA
InpValAt20mA
FilterTC
Table 1‐1. ProgramModeFunctions(continued)
Option
#
Function Parameters/Description Purpose