KurzBSeriesOperationsGuide 1–13
BSeriesFlowMeterModes & Menus
16 REMOTECF RemoteCorrectionFactorData
RemoteCF
ON OFF
Enter#ofRemoteCFDataPointsn
EnterExtInput.RemCFD1
EnterExtInput.RemCFX1
:
EnterExtInput.RemCFDn
EnterExtInput.RemCFXn
Thisallowsyouto
correctthemeter
readinginreference
tootherdata.
17 CALCURVE GasCalibrationCurveorGasConcentration
Selection(optional)
CurveSel
Mode
MANUALSELECT
EXTINPU TLEV
IfMANUALSELECT
CalCurve#
Theselection
dependsonthe
meterconfiguration.
18 DATALOG DataLoggin gSetup
EnableDataLog
ON OFF
IfON
LogIntervalsec
Setsupperiodic
loggingofmeasured
flowandtotalflow
totheUSBport.
19 MODBUSCOM ModbusCommunicationSetup
DevModbusAddr
Modbus
Mode
MODBUSRTU MODBUSASCII
ModbusBaudRate
9600 38400
14400 57600
19200
RegisterOrder
BYTE#1234 BYTE#3412
Allowsyoutosetup
upModbus
communications
andspecifyeither
RTUorASCII
commands.The
baudratewill
typicallybe9600.
20 EXTAINPUT ExternalInputSetup
ExtInputUsage
CALDATASW PIDEXT.REF(optional)
VMREFERENCE REMOTECF
PURGECOMMAND
(optional)
VMREFERENCE
ScaleUnit
InpValAt4mA
InpValAt20mA
FilterTC
Table 11. ProgramModeFunctions(continued)
Option
#
Function Parameters/Description Purpose

Table of Contents

Other manuals for Kurz B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kurz B Series and is the answer not in the manual?

Kurz B Series Specifications

General IconGeneral
BrandKurz
ModelB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals