82 Flowmaster User’s and Service Manual | LBL-000990 [B] Appendix A
compared to the previous software version due to
changed/improved analysis techniques or changed
medical opinion/insight.
Never use any detergents, thinner, benzine or other volatile
chemicals while cleaning the flowsensor to prevent
damage to the system.
Do not use excessive amounts of water for cleaning the
flowsensor to prevent damaging the electronics.
Never allow water to get inside the flowsensor to prevent
damage to the system.
Do not use chloride-based solutions for cleaning the flow
stand. Check that the solution does not affect stainless
steel (AISI304L). Clean the flow stand if it was
contaminated with urine.
Repair (Laborie qualified personnel only)
Be sure to perform a safety test in accordance to IEC
60601-1 after modules/devices have been replaced or
opened.
Before repair make sure that the flowsensor is switched
off and disconnected from the mains.
Before repair always take antistatic precautions.
Although precautions have been taken in this apparatus to prevent
malfunctions due to abnormal environmental conditions and/or abnormal
handling, such malfunction can occur in some circumstances.