170 Chapter 7
Controlling the OBR Remotely
SCAN
Tells the OBR to execute an optical Scan based on the
configured system parameters.
124
SYST:ACQ? Queries whether data has been acquired (1) or not (0). 131
SYST:ALIG
Aligns the optics
126
SYST:ALIG? Queries the OBR if the optics are aligned (1) or not (0). 127
SYST:CAL
Calibrates the system.
127
SYST:CAL? Queries the system calibration status. Returns ā€œ1ā€ if the
system is calibrated successfully or ā€œ0ā€ if it is not.
127
SYST:CALF {0,
filename}
Sets the name of the calibration file. If the filename is not
specified, this command sets calibration file to the default
filename, obrCal_refl.
128
SYST:CALF? 0 Retrieves the name of the calibration file. 129
SYST:COPY
{source, destination}
Copies data from the source trace (A-E) to the destination
trace (A-E).
131
SYST:ERR?
Returns a numeric error code. Zero (0) indicates no error; i.e.
the last command or query completed successfully. Any non-
zero number indicates that an error has occurred. A
description of the error may be retrieved with the query
SYST:ERRD?
126
SYST:ERRD? Retrieves the detailed error description for the most recent
remote operation.
126
SYST:GET? {trace} Queries whether data is loaded in the specified trace (A-E).
131
SYST:LASE {0|1} Turns the laser on (1) or off (0). 125
SYST:LASE? Queries if the laser is on (1) or off (0). 126
SYST:LAST? Queries the last remote command or query processed 132
SYST:LOAD
{
"filename.obr",
trace}
Loads the external file specified in quotes as a source of data
reference for the measurement.
129
SYST:ON? Queries whether the OBR is on (1) or off (0). 125
SYST:RDY? Queries if the OBR is ready to scan (1) or not (0). 126
Table 7-1: Alphabetical Summary of GPIB and TCP/IP Commands