MINI Hawk High Performance Imager User’s Manual 6-21
Symbologies
Large Intercharacter Gap (Codabar)
Fixed Symbol Length Status (Codabar)
Symbol Length (Codabar)
Definition: When disabled, the spaces between characters, or the “intercharacter
gap”, are ignored during the decode process.
Note: If the intercharacter space is large enough to be considered a margin,
the symbol will not decode, regardless of this parameter’s setting.
Serial Cmd: <K471,status,start/stop match,start/stop output,large intercharacter
gap,fixed symbol length status,symbol length,check character type,check
character output>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Definition: When disabled, the imager will accept any Codabar symbol provided it
doesn’t exceed the system’s maximum capabilities.
When enabled, the imager will reject any Codabar symbol that doesn’t
match the fixed length.
Serial Cmd:
<
K471,
status,start/stop match,start/stop output,large intercharacter gap,
fixed
symbol length status
,symbol length,check character type,check character
output>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Definition: This is the value against which all Codabar symbol lengths will be compared.
Serial Cmd: <K471,status,start/stop match,start/stop output,large intercharacter gap,fixed
symbol length status,symbol length,check character type,check character
output>
Default: 10
Options: 1 to 64