Monitoring NIBP
9-6
9.7 Changing NIBP Settings
By selecting the NIBP parameter window, you can enter the [NIBP Setup] menu.
9.7.1 Choosing NIBP Alarm Source
You can monitor for alarm conditions in systolic, diastolic and measure pressure, either
singly or parallel. In the [NIBP Setup], select [Alm Source] and choose from:
îš„ [Sys]: alarms are given only when the systolic pressure violates the alarm limits.
îš„ [Dia]: alarms are given only when the diastolic pressure violates the alarm limits.
îš„ [Mean]: alarms are given only when the mean pressure violates the alarm limits.
îš„ [Sys&Mean]: alarms are given when the systolic or mean pressure violates the alarm
limits.
îš„ [Mean&Dia]: alarms are given when the mean or diastolic pressure violates the alarm
limits.
îš„ [Sys&Dia]: alarms are given when the systolic or diastolic pressure violates the alarm
limits.
îš„ [All]: alarms are given when the systolic or mean or diastolic pressure violates the alarm
limits.
9.7.2 Displaying NIBP Measurements
You can display one set or multiple sets of NIBP measurements simultaneously. In the [NIBP
Setup] menu, select [Display NIBP] and toggle between [Single-Group] and
[Multi-Group]:
îš„ [Single-Group]: only the latest set of NIBP measurements is displayed in the NIBP
parameter window.
îš„ [Multi-Group]: multiple sets of most recent NIBP measurements are displayed below
the waveform area.

Table of Contents

Other manuals for Mindray BeneView T8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T8 and is the answer not in the manual?

Mindray BeneView T8 Specifications

General IconGeneral
BrandMindray
ModelBeneView T8
CategoryMonitor
LanguageEnglish

Related product manuals