Factory Defaults
C-3
Arrh. Alarm Setup Alarm Alm Lev Alm Rec
Asystole On High Off
VFib/VTac On High Off
R on T On Med Off
VT>2 On Med Off
Couplet On Med Off
PVC On Med Off
Bigeminy On Med Off
Trigeminy On Med Off
Tachy On Med Off
Brady On Med Off
PNP On Med Off
PNC On Med Off
Missed Beat On Med Off
Waveform X * Adult Pediatric Neonate
HR Source II
Gain X1
Filter Diagnostic
Sweep 25 mm/s
Cascade Off
*: X represents (lead) I, II, III, aVR, aVL, aVF, V, V1, V2, V3, V4, V5 or V6
C.5 Resp Setup
Resp Setup Adult Pediatric Neonate
Alarm On
Alm Lev Med
Alm Rec Off
Resp High 30 30 100
Resp Low 8 8 30
Apnea Time 20 s
Detection Mode Auto
Resp Waveform