A-1
A Product Specifications
A.1 Monitor Safety Specifications
A.1.1 Classifications
The patient monitor is classified, according to IEC60601-1:
Components
Type of protection
against electrical
shock
Degree of protection
against electrical
shock
Degree of protection
against harmful
ingress of water
Degree of protection
against hazards of
explosion
Mode of operation
Main unit Not marked
Secondary display
I
Not marked
MPM
IBP module
C.O. module
ICG module
CF
BIS module
AG module
CO
2
module
RM module
BF
SMR
NA
Not marked
Ordinary Not suitable Continuous
I: Class I equipment
BF: Type BF applied part
CF: Type CF applied part
NA: Not applicable
Ordinary: Ordinary equipment (enclosed equipment without protection against ingress
of water)
Not suitable: Equipment not suitable for use in the presence of a flammable anesthetic
mixture with air with oxygen or nitrous oxide.