Monitoring ECG/Resp
7-17
7.8 12-Lead ECG Monitoring
7.8.1 Entering the 12-lead ECG Analysis Program
1. Refer to the section
7.3.3 ECG Lead Placements for placing the electrodes.
2. In the [ECG Setup] menu, select [Lead Set][12-Lead], and select [ECG Display]
[12-Lead].
The ECG numerics are displayed above the ECG wave. There are totally 12 ECG waves and
1 rhythm wave displayed on the screen. Before entering the 12-lead ECG analysis program,
the rhythm lead is the lead from which HR is derived. The ST numerics are displayed in three
groups:
ST Ant (anterior): V1, V2, V3, V4
ST Inf (inferior): II, III, aVF, (aVR)
ST Lat (lateral): I, aVL, V5, V6