Calculations
20-9
20.5 Hemodynamic Calculations
20.5.1 Performing Calculations
1. By selecting [Main Menu]→[Calculations >>]→[Hemodynamic >>], or selecting
[Calculations] QuickKey→[Hemodynamic >>], you can access the following
window.
2. Enter values for calculation.
îš‹ For a patient who is being monitored, [HR], [Art mean], [PA mean] and [CVP]
are automatically taken from the currently measured values. If you just have
performed C.O. measurements, [C.O.] is the average of multiple thermodilution
measurements. [Height] and [Weight] are the patient’s height and weight you
have entered. If the monitor does not provide these values, their fields appear
blank.
îš‹ For a patient who is not being monitored, confirm the values you have entered.
3. Select the [Calculate] button. The system performs a calculation per the current
settings and displays the calculated values.