Factory Defaults
C-5
C.7 NIBP Setup
NIBP Setup Adult Pediatric Neonate
Alarm On
Alm Lev Med
Alm Rec Off
Alm Source All
Sys High 160 120 90
Sys Low 90 70 40
Mean High 110 90 70
Mean Low 60 50 25
Dia High 90 70 60
Dia Low 50 40 20
Display NIBP Single-group
Press. Unit mmHg
Interval Manual
Cuff Press. 80 60 40
C.8 Temp Setup
Temp Setup Adult Pediatric Neonate
Alarm On
Alm Lev Med
Alm Rec Off
T1 High 39.0 39.0 39.0
T1 Low 36.0 36.0 36.0
T2 High 39.0 39.0 39.0
T2 Low 36.0 36.0 36.0
TD High 2.0 2.0 2.0
Unit ºC