EasyManua.ls Logo

Mindray BeneView T8 - Safety Information

Mindray BeneView T8
290 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Monitoring BIS
16-2
16.2 Safety Information
For patients with neurological disorders, patients taking psychoactive medication, and
children below the age of 1 year, BIS values should be interpreted cautiously.
WARNING
z The conductive parts of sensors and connectors should not come into contact with
other conductive parts, including earth.
z To reduce the hazard of burns in the high-frequency surgical neutral electrode
connection, the BIS sensor should not be located between the surgical site and the
electro-surgical unit return electrode.
z The BIS sensor must not be located between defibrillator pads when a defibrillator
is used on a patient connected to the patient monitor.

Table of Contents

Other manuals for Mindray BeneView T8

Related product manuals