EasyManua.ls Logo

Mindray BeneView T8 - C.15 RM Setup

Mindray BeneView T8
290 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Factory Defaults
C-12
C.15 RM Setup
Factory default settings
RM Setup
Adult Pediatric Neonate
Alarm On
Alm Lev Med
Alm Rec Off
Apnea Time 20 s
TV / MV TV
Sensor Type Disposable
Ventilation Mode Spontaneous
Alarm Limits Setup
RR High 30 30 100
RR Low 8 8 30
PEEP High 10
PEEP Low 0
PIP High 40
PIP Low 1
MVe High 30.0 30.0 10.0
MVe Low 2.0 2.0 0.5
RM Waveform
Sweep 25 mm/s

Table of Contents