22 Other Functions
22.1 Marking Events
During patient monitoring, some events may exert effects on the patient and as a result
change the waveforms or numerics displayed on the monitor. To help analysing the
waveforms or numerics at that time, you can mark these events.
To mark an event:
1. Select [Mark Event] QuickKey, or [Main Menu]→[Mark Event >>].
2. Select [Event A] and the symbol [@] is displayed on the option.
When reviewing graphic trends, tabular trends or full-disclosure waveforms, the event
symbols such as A, B, C and D are displayed at the time the event is triggered.
22.2 Analog Output
The patient monitor provides analog output signals to accessory equipment via the Auxiliary
Output Connector on the rear of the monitor. To obtain analog output signals, connect the
accessory equipment such as an oscillograph, etc. to the monitor and then follow this
procedure:
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. Select [Others >>] to open the [Others] menu.
3. Select [Auxiliary Output] and then select [Analog Output].
4. Select [Main Menu]→[Analog Output Setup>>].
5. Select [Analog Output] and then select [On].
6. Select [Waveform] and then select a waveform you want to output.
NOTE
z The analog output feature is seldom applied in clinical applications. You can
contact your service personnel for more details.

Table of Contents

Other manuals for Mindray BeneView T8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T8 and is the answer not in the manual?

Mindray BeneView T8 Specifications

General IconGeneral
BrandMindray
ModelBeneView T8
CategoryMonitor
LanguageEnglish

Related product manuals