Factory Defaults
C-6
C.9 IBP Setup
IBP Setup Adult Pediatric Neonate
Alarm On
Alm Lev Med
Alm Rec Off
Alm Source All
Systolic/Diastolic (Mean) (mmHg)
Alarm Limits
Adult Pediatric Adult
High Limit 160/90 (110) 120/70 (90) 90/60 (70) Art, Ao, FAP,
BAP, UAP
Low Limit 90/50 (70) 70/40 (50) 55/20 (35)
High Limit 35/16 (20) 60/4 (26) 60/4 (26) PA
Low Limit 10/0 (0) 24/-4 (12) 24/-4 (12)
High Limit 160/90 (110) 120/70 (90) 90/60 (70) Custom
pressures, P1 to
P8
Low Limit 90/50 (70) 70/40 (50) 55/20 (35)
Mean (cmH
2
O)
Alarm Limits
Adult Pediatric Adult
High Limit 7.4 2.9 2.9 CVP, L A P, R A P,
ICP, UVP
Low Limit 0.0 0.0 0.0
IBP Waveform
Sweep 25 mm/s
Filter No Filter