Factory Defaults
C-4
Lead II
Sweep 12.5 mm/s
Gain X2
Detection Mode Auto
C.6 SpO
2
Setup
SpO
2
Setup Adult Pediatric Neonate
Alarm On
Alm Lev Med
Alm Rec Off
SpO2 High 100 100 95
SpO2 Low 90 90 80
Desat Limit 80 80 70
Sensitivity Mindray) Med
Sensitivity (Masimo) Normal
Averaging (Masimo, Nellcor) 8 s
NIBP Simul Off
Sat-Seconds (Nellcor) 0 s
PR Setup Adult Pediatric Neonate
PR On
Alarm On
Alm Lev Med
Alm Rec Off
PR High 120 160 200
PR Low 50 75 100
Alm Source Auto
Beat Vol 2
Pleth Waveform
Sweep 25 mm/s