Product Specifications
A-2
A.1.2 Environmental Specifications
Main unit, MPM, IBP module, C.O. module, Recorder
Item Operating conditions Storage conditions
Temperature (ºC) 5 to 40 -20 to 60
Relative humidity (noncondensing) 15 to 95% 10 to 95%
Barometric (kPa) 70.0 to 106.0 22.0 to 107.4
Microstream CO
2
module
Item Operating conditions Storage conditions
Temperature (ºC) 0 to 40 -20 to 60
Relative humidity (noncondensing) 15 to 95% 10 to 95%
Barometric (kPa) 430 to 795 430 to 795
Sidestream CO
2
module
Item Operating conditions Storage conditions
Temperature (ºC) 5 to 35 -20 to 60
Relative humidity (noncondensing) 15 to 95% 15 to 95%
Barometric (kPa) 428 to 790 428 to 790
Mainstream CO
2
module
Item Operating conditions Storage conditions
Temperature (ºC) 10 to 40 -10 to 50
Relative humidity (noncondensing) 10 to 90% 0 to 90%
Barometric (kPa) 400 to 850 400 to 850
AG module
Item Operating conditions Storage conditions
Temperature (ºC) 10 to 40 -20 to 70
Relative humidity (noncondensing) 15 to 95% 10 to 95%
Barometric (kPa) 525 to 900 525 to 900

Table of Contents

Other manuals for Mindray BeneView T8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T8 and is the answer not in the manual?

Mindray BeneView T8 Specifications

General IconGeneral
BrandMindray
ModelBeneView T8
CategoryMonitor
LanguageEnglish

Related product manuals