Monitoring Cardiac Output
12-4
4. Select the C.O. parameter window to enter the [C.O. Setup] menu. In the [C.O. Setup]
menu:
Check that the correct computation constant is entered. To change the
computation constant, select [Comp. Const] and then enter the correct value.
When a new catheter is used, the computation constant should be adjusted in
accordance with the manufacturer’s instructions for use.
Switch [Auto IT] on and the injectate temperature is automatically obtained.
When [Auto IT]] is switched off, you can enter the injectate temperature in the
[Manual IT] field.
5. Select [C.O. Measure] to enter the C.O. measurements window.
Currentl
y
measured numeric
Currentl
measured C.O. curve
Prom
p
t messa
g
e area
measurement windows
Averaged values
Buttons