Product Specifications
A-16
A.5.6 IBP
Standards Meet standard of EN60601-2-34/IEC60601-2-34.
Technique Direct invasive measurement
IBP
ART 0 to 300 mmHg
PA -6 to 120 mmHg
CVP/RAP/LAP/ICP -10 to 40 mmHg
Measurement range
P1/P2 -50 to 300 mmHg
Resolution 1 mmHg
Accuracy ±2% or ±1mmHg, whichever is greater
Refreshing rate 1s
Pressure transducer
Excitement voltage 5 VDC, ±2%
Sensitivity 5 uV/V/mmHg
Impedance range 300 to 3000Ω
A.5.7 C.O.
Measurement method Thermodilution method
C.O.: 0.1 to 20 L/min
BT:
23 to 43°C
Measurement range
TI:
0 to 27°C
Resolution
C.O.:
BT, TI:
0.1 L/min
0.1°C
Accuracy
C.O.:
BT, TI:
±5% or ±0.1 L /min
±0.1°C (without sensor)
Alarm range BT:
23 to 43°C