EasyManua.ls Logo

Mindray BeneView T8 - Page 268

Mindray BeneView T8
290 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Factory Defaults
C-8
C.11 CO
2
Setup
Factory default settings
CO
2
Setup
Adult Pediatric Adult
Alarm On
Alm Lev Med
Alm Rec Off
Humidity Compen
(side/microstream)
Wet
Press. Unit mmHg
Max Hold (microstream) 20 s
Max Hold (mainstream) 10 s
Operating Mode (side/mainstream) Standby
Operating Mode (microstream) Measure
O2 Compen (side/mainstream) Off
N2O Compen (sidestream) Off
Des Compen (sidestream) Off
Balance Gas (mainstream) Room Air
Auto Standby (min) (microstream) 0
EtCO2 High 50 50 45
EtCO2 Low 15 20 30
FiCO2 High 4 4 4
awRR High 30 30 100
awRR High 8 8 30
Apnea Time 20 s
CO
2
Waveform
Wave Type Draw
Sweep 25 mm/s

Table of Contents

Other manuals for Mindray BeneView T8

Related product manuals