Factory Defaults
C-8
C.11 CO
2
Setup
Factory default settings
CO
2
Setup
Adult Pediatric Adult
Alarm On
Alm Lev Med
Alm Rec Off
Humidity Compen
(side/microstream)
Wet
Press. Unit mmHg
Max Hold (microstream) 20 s
Max Hold (mainstream) 10 s
Operating Mode (side/mainstream) Standby
Operating Mode (microstream) Measure
O2 Compen (side/mainstream) Off
N2O Compen (sidestream) Off
Des Compen (sidestream) Off
Balance Gas (mainstream) Room Air
Auto Standby (min) (microstream) 0
EtCO2 High 50 50 45
EtCO2 Low 15 20 30
FiCO2 High 4 4 4
awRR High 30 30 100
awRR High 8 8 30
Apnea Time 20 s
CO
2
Waveform
Wave Type Draw
Sweep 25 mm/s