Alarm Messages
D-7
Measure
ment
Alarm message A B Cause and solution
AG Change Watertrap √ × Wait until the change is completed.
AG Watertrap Type Wrong √ × Make sure that a correct watertrap has been
used.
O2 Accuracy Unspecified × √
N2O Accuracy
Unspecified
√ ×
CO2 Accuracy
Unspecified
√ ×
Enf Accuracy Unspecified √ ×
Iso Accuracy Unspecified √ ×
Sev Accuracy Unspecified √ ×
Hal Accuracy Unspecified √ ×
Des Accuracy Unspecified √ ×
RR Accuracy Unspecified √ ×
The measured value has exceeded the specified
accuracy range.
O2 Sensor Err √ ×
Galvanic O2 Sensor Err √ ×
AG Hardware Err √ ×
An error occurred to the module. Re-plug the
module or restart the monitor.
AG Airway Occluded √ × Check the airway and remove the occlusion.
AG Zero Failed √ × Re-plug the module or restart the monitor, and
then perform a zero calibration again.
AG Oxima Depletion Err √ ×
AG Oxima Depletion
Warn
√ ×
Contact your service personnel and replace the
O2 battery.
AG Accuracy Err √ × The measured value has exceeded the specified
accuracy range.
RM No Sensor √ ×
RM Sensor Reversed × ×
Check and reconnect the sensor.
Zero Failed × × Perform a zero calibration again.
RM
RM Power Err × √ There is a problem with the power supply.
Re-plug the module or restart the monitor.
BIS High Imped. √ ×
BIS Sensor Off √ ×
Check and reconnect the BIS sensor.
BIS DSC Err × × An error occurred to the DSC during receiving
signals. Check the DSC.
BIS
BIS DSC Malf × × The DSC automatically shuts down as a result of
malfunction. Check the DSC.