7-14 Operator’s Manual – OPTI CCA-TS
7 DIAGNOSTICS AND TROUBLESHOOTING
7.2.2 Checking System Temperatures
From the <Ready> display, press <System
Manager -> Diagnostics>.
The <Temperature> option lets you check
the various system temperatures: <Top Plate>,
<Bottom Plate> and <Ambient> (Fig. 7-3).
NOTE: If top or bottom plate temperatures are
out of range, the temperature display will
change to red.
Fig. 7-3 Temperature
7.2.3 Checking Gas Pressure
Fig. 7-4 Gas Pressure
The <Gas Level> option allows you to check the
percent remaining of the gas bottle (Fig. 7-4).
With a new gas bottle in place, the pressure should
be approx. 99%, with the bottle removed, the
pressure should be 00%.
From the <Ready> display, press <System
Manager> and <Diagnostics>.
• To install a new gas bottle, press
.
• Scan the bar code located on the side of the gas
bottle to install a new gas bottle (see Section
6.5.1 “Changing Gas Bottle”) .
• Press
to initiate a gas purge or press
to cancel this function and return to
the <Diagnostics> screen.