D-2 Operator’s Manual – OPTI CCA-TS
APPENDIX D - REPORT FORMATS
SRC Measurement Report
(E-Ca example)
----------------------------
OPTI Medical OPTI CCA-TS
SRC Measurement
DD-MMM-YY HH:MM
Level: 3
SRC ID:XXXXX Exp: MMMYY
RESULT LIMITS OK?
pH 7.601 7.580-7.620 OK
PCO2 20.0 18.0-22.0 OK
PO2 170.0 167.0-173.0 OK
Na+ 165.1 142-148 OK
K+ 7.00 6.70-7.30 OK
Ca++ 0.70 0.60-0.80 OK
tHb 8.5 6.5- 9.5 OK
SO2 98.6 96.0-100.0 OK
SRC Test Result: PASS
Operator ID:123456789012
S/N:123456
----------------------------