Operator’s Manual – OPTI CCA-TS 6-1
6 MAINTENANCE
6 MAINTENANCE
6.1 Daily Maintenance
No daily maintenance is required for the OPTI CCA-TS system.
6.2 Weekly Maintenance
Once a week, the Sample Measurement Chamber (SMC) must be cleaned. Open the top cover
and clean the optics surface as well as the underside of the SMC cover with a lint-free cloth,
dampened with a dilute alcohol or dilute bleach cleaner as needed. Be sure to remove all blood
residue. A cotton swab may be used for cleaning the smaller parts of the SMC.
6.3 Quarterly Maintenance – Performing tHb Calibration
6.3.1 Performing HbCal - tHb-Calibrator
1. In the <Ready> display, select <QC
Manager>.
2. In the <QC Manager> menu, select
<tHb-Calibrator> (Fig. 6-1).