Operator’s Manual – OPTI CCA-TS 7-15
7 DIAGNOSTICS AND TROUBLESHOOTING
7.2.4 Checking the LEDs
This menu displays the current barometric pressure.
From the <Ready> display, press
<System Manager -> Diagnostics>.
• The <Miscellaneous> section will show the
current barometric pressure (Fig. 7-6).
• If the barometric pressure requires adjustment,
refer to Setup, Section 3.3.3.2 “Entering the
Barometric Pressure” for setting the barometer.
Fig. 7-6 Barometric Pressure
7.2.5 Verifying Barometric Pressure
Fig. 7-5 LEDs
The purpose of this test is to check proper
functioning of the LEDs. This test should be
performed only by trained service personnel.
From the <Ready> display, press <System
Manager -> Diagnostics>.
The following information is displayed in the
<LEDs> section (Fig. 7-5):
• <Front>, <Rear>, <Ion> - uid light gates.
• <Cassette Misseat> detector (located in
cover)
• <Cassette Detect> sensor
• <SMC Cover> - this test function indicates
whether the SMC cover is closed or open.
• Press
to return to the <System>
screen.