3-4 Operator’s Manual – OPTI CCA-TS
3 CUSTOMIZATION
3.3.1.2 Setting up the Quality Control Material Lot and Level
1. From the <Ready> display, select
<System Manager> and <Setup>.
2. Enter the password if this security function has
been activated under <Setup>.
3. Select <Control> (Fig. 3-9).
4. Scan the 36-digit bar code marked SCAN A
for the applicable level supplied with OPTI
CHECK or OPTI CHECK PLUS.
5. Scan the second 36-digit bar code marked
SCAN B supplied with the same OPTI CHECK
or OPTI CHECK PLUS control lot. These two
bar codes contain all necessary lot information
for each level, and may be conrmed in the
subsequent screens.
6. When using OPTI CHECK PLUS, scan the
third 36-digit bar code marked SCAN C for the
applicable level.
NOTE: If bar code is damaged and won’t scan,
press <Manual Entry> on the <Scan
Barcode> screen and enter the barcode
information manually.
Fig. 3-9 Select Control
NOTE: If previous SRC data exists, the unit will
prompt the user to either print and/or delete
this data from the database.
NOTE: If no previous QC data exists in the
database, the print and delete display
screens will be bypassed.
6. Press to obtain a printout of the old
SRC database.
7. Press to delete the old SRC database.
To continue quality control (SRC) programming,
repeat the above procedure for all levels of SRCs.
Print Old SRC Data?
Setup QC . . .
Yes
No
Print Old SRC Data?
Setup QC . . .
Yes
No
Delete Old SRC Data?
Setup QC . . .
Yes
No
Delete Old SRC Data?
Setup QC . . .
Yes
No