D-4 Operator’s Manual – OPTI CCA-TS
APPENDIX D - REPORT FORMATS
Controls Measurement Report
(E-Ca example)
----------------------------
OPTI Medical OPTI CCA-TS
Controls Measurement
DD-MMM-YY HH:MM
Level: 2 OPTI-check
Sample No.:2345
QCLot:1234 Exp:MMMYYYY
RESULT LIMITS OK?
pH 7.551 7.48-7.56 OK
PCO2 69.8 65-75 OK
PO2 99.8 96-105 OK
Na+ 144.2 142-148 OK
K+ 4.46 4.0-4.8 OK
Ca++ 1.10 1.0-1.2 OK
tHb 14.7 14.0-15.3 OK
SO2 98.3 96-99 OK
Control Test Result: PASS
Store to Database: YES
Barometer: 744.5 mmHg
Operator ID:12345678901
S/N:1234 Lot:123456
---------------------------